Home » How to Plan an Accelerated Aging Study Under ASTM F1980

Your product is ready to launch, and it needs a three-year shelf-life claim. You don’t have three years to wait. That’s the problem accelerated aging solves. Heat speeds up the chemical changes that happen to packaging over time, so months or years of storage can be simulated in weeks.

We’ve written about how accelerated aging predicts real-world performance. This post is the planning guide. It covers the decisions you’ll make before samples go into the chamber, and the math behind the test time.

Who needs an accelerated aging study?

Any company that has to back up a shelf-life or expiration claim. That includes:

  • Medical device makers proving sterile barrier integrity over the product’s life
  • Pharmaceutical companies testing blister packs and bottles
  • Food and beverage brands checking shelf stability
  • Consumer goods makers checking labels, seals and product performance

Step 1: Know the standard

The go-to guide is ASTM F1980, the standard guide for accelerated aging of sterile barrier systems. Medical device teams use it most, but other industries borrow its method. Before planning, confirm the ASTM requirements your customer or regulator expects. We apply F1980 alongside FDA and ISO guidance.

One point trips people up: under F1980, accelerated aging supports a shelf-life claim until real-time data is available. Plan to run real-time aging samples alongside your accelerated set.

Step 2: Pick the aging temperature

Higher temperatures mean shorter tests. But too much heat can cause damage that would never happen on a warehouse shelf. Some plastics soften or warp well before they “age.” F1980 generally advises keeping aging temperatures below 60°C (140°F). Choose a temperature that speeds things up without changing how your materials behave.

Step 3: Calculate the test time

F1980 uses a simplified form of the Arrhenius equation. It assumes the rate of aging roughly doubles for every 10°C rise in temperature. That doubling factor is called Q10, and a Q10 of 2 is the common conservative choice.

AAF = Q_{10}^{(T_{AA} - T_{RT})/10}

AAF is the accelerated aging factor. T_AA is the aging temperature and T_RT is normal room temperature, both in °C. Divide the shelf life you want to claim by the AAF to get the chamber time.

Here is a worked example with a 55°C chamber, 25°C room temperature and a Q10 of 2:

Shelf-life claim

AAF

Accelerated aging test time

1 year

2³ = 8

about 46 days

2 years

8

about 91 days

3 years

8

about 137 days

These numbers are an illustration. Your final accelerated aging test plan depends on your materials, humidity needs and the claim you want to make.

Step 4: Decide on humidity

Temperature isn’t the only factor. Paper, adhesives and some films react to moisture. If your product will see humid storage, build humidity into the plan. Just don’t let extreme moisture create failures that wouldn’t happen in real life. Our post on controlled conditioning covers how conditioning works.

Step 5: Plan the testing after aging

Aging alone proves nothing. The aged packages have to be tested to show they still work. Most plans look for:

  • Seal strength and seal degradation
  • Barrier failures (moisture, oxygen)
  • Changes in material strength or flexibility
  • Ink fading or smearing on labels
  • Package deformation

Many protocols also run aged packages through a shipping simulation, such as the ASTM D4169 distribution test. That shows the package can survive shipping at the end of its shelf life, not just the start.

Accelerated aging planning checklist

  • Shelf-life claim defined (for example, 2 years)
  • Standard confirmed (ASTM F1980 or customer spec)
  • Aging temperature chosen and below material limits
  • Q10 value and AAF calculated
  • Humidity requirements set
  • Real-time aging samples set aside
  • Post-aging tests and pass criteria written down
  • Enough samples for every time point and test

Why work with an independent lab

Good accelerated aging package testing depends on tight chamber control. Our calibrated conditioning chambers hold temperature and humidity steady, so results are consistent and repeatable. And an independent, third-party report carries weight with customers and regulators.

A packaging lab near St. Louis

Pro-Pack Testing Laboratory has been testing packages since 1982. We’re at 2385 Amann Drive in Belleville, Illinois, about 20 minutes from downtown St. Louis. We serve medical device, pharmaceutical, food and beverage, and consumer goods companies.

Start your shelf-life study

Tell us your shelf-life claim and we’ll help you build the test plan. Call Pro-Pack at 618-277-1160 or contact us to request a quote.

Get Started with ProPack Test Lab:

2385 Amann Drive
Belleville, Illinois 62220
Phone: 618-277-1160
Fax: 618-277-1163